Sterilization review for the CRCST

Steam is the default method: a gravity-displacement cycle for wrapped items typically runs 30 minutes at 121 C (250 F), while a dynamic-air-removal (prevacuum) cycle runs 4 minutes at 132 C (270 F).

Updated 2026-09-23 · 2 sources · By the HealthCert Prep team

Educational exam prep, not medical advice. This page teaches what certification exams test. Follow your facility’s policies, your supervising clinician and the manufacturer’s instructions in practice. If you are a patient with a health question, talk to your clinician or see MedlinePlus.

Typical steam cycles for wrapped items (CDC)

Minimum cycle examples from the CDC Guideline for Disinfection and Sterilization. Always follow the device manufacturer's instructions for use (IFU); AAMI ST79 is the industry standard for steam sterilization.
Sterilizer typeTemperatureExposure time
Gravity displacement121 C (250 F)30 minutes
Gravity displacement132 C (270 F)15 minutes
Dynamic air removal (prevacuum)132 C (270 F)4 minutes

Monitoring

ToolWhat it tells you
Bowie-Dick testAir removal in dynamic-air-removal sterilizers; run daily in an empty chamber, usually the first cycle
Biological indicator (BI)Whether spores were killed: Geobacillus stearothermophilus for steam and hydrogen peroxide; Bacillus atrophaeus for EtO and dry heat
Chemical indicators (ISO 11140-1 types 1-6)Type 1 process indicators (tape); Type 2 specific tests (Bowie-Dick); Types 3-6 single-variable, multi-variable, integrating and emulating
Mechanical monitoringCycle printouts/displays of time, temperature and pressure for every load

Low-temperature methods

MethodKey points
Ethylene oxide (EtO)For heat- and moisture-sensitive items; requires aeration to remove toxic residue
Hydrogen peroxide gas plasma / vaporLow temperature, no toxic residue; cellulose (paper, linen) and liquids are not compatible

Spaulding classification

CategoryContactsMinimum processing
CriticalSterile tissue or the vascular systemSterilization
SemicriticalMucous membranes or non-intact skinHigh-level disinfection (minimum)
NoncriticalIntact skinLow-level disinfection

Test yourself on the CRCST

Take the free practice test and get a score report by topic in a few minutes.

Start the free CRCST practice test

Study every exam with one subscription

All-access is $29/month (cancel anytime) or $59 for 90 days with no auto-renew. Full original question banks, timed practice exams and score reports by domain.

Get all-access - $29/mo

Frequently asked questions

What is a Bowie-Dick test?

A daily air-removal test for dynamic-air-removal steam sterilizers, run in an empty chamber.

Which organism is used in steam biological indicators?

Geobacillus stearothermophilus spores.

What is the Spaulding classification?

A system that ranks devices as critical, semicritical or noncritical by infection risk and sets the minimum processing level for each.